Abstract

This dissertation investigated AISI 1020 low-carbon steel as an alternative vacuum chamber material to conventional stainless steel for ultra-high vacuum (UHV) and extreme-high vacuum (XHV) applications. After a 400 °C/48 h bake, AISI 1020 tube chambers achieved a hydrogen outgassing rate of 2.4 × 10⁻¹⁶ Torr·L·s⁻¹·cm⁻², approximately 2,300 times lower than the prebaked 316L stainless-steel comparator, among the lowest hydrogen outgassing rates ever reported for an uncoated metallic vacuum chamber. Bare and magnetite-coated AISI 1020 chambers were then compared using throughput and rate-of-rise methods. The magnetite coating yielded 5× lower water outgassing at room temperature, but this advantage disappeared after 80 °C baking. After full thermal conditioning (400 °C/48 h prebake followed by 150 °C/96 h and 200 °C/110 h), bare steel achieved 25× lower hydrogen outgassing than the magnetite-coated chamber (9.6 × 10⁻¹⁶ Torr·L·s⁻¹·cm⁻²) and >99% H₂ purity with carbon species below RGA detection. Monte Carlo molecular flow simulations of a CEBAF photogun beamline (96 scenarios) showed that replacing 304L stainless steel with AISI 1020 reduces equilibrium H₂ pressure by a factor of 833; a single 304L electrode contributes 98.8% of the gas load despite occupying only 9.1% of the internal surface area. A 500-m Einstein Telescope beampipe screening showed that corrugated bellows contribute 18% of the gas load from only 0.7% of the surface area. A five-model adsorption isotherm framework applied to 22 pumpdown datasets (164 fits with AR(1)-GLS correction) established that the experimental protocol, not the material, controls isotherm identifiability: Dubinin–Radushkevich wins isothermal pipe pumpdowns; Langmuir wins thermally dominated chamber bakes. Cross-dataset joint fitting of the AISI 1020 pipe pumpdowns yielded an H₂ diffusion activation energy Ed = 7.24 ± 1.28 kcal·mol⁻¹, consistent with trap dominated diffusion in commercial low-carbon steels. Two companion innovations were developed: a Variable Conductance Device (VCD, patent pending IDF-00723) for XHV outgassing measurement, and VacuumDesignerPro (VDP), a MATLAB-based design tool validated against LIGO benchmarks.

Rights

In Copyright. URI: http://rightsstatements.org/vocab/InC/1.0/ This Item is protected by copyright and/or related rights. You are free to use this Item in any way that is permitted by the copyright and related rights legislation that applies to your use. For other uses you need to obtain permission from the rights-holder(s).

DOI

https://doi.org/10.25777/jac2-9b35

ISBN

9798197810632

ORCID

0000-0002-0415-2795

Share

COinS